Journal article

Evaluation of geometrical sensitivity for respiratory motion gating by GATE and NCAT simulation.

Jianfeng He, Graeme J O'Keefe, Gareth Jones, Tim Saunder, Sylvia J Gong, Moshi Geso, Andrew M Scott

Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC) | Published : 2007

Abstract

Respiratory motion artifacts can be a significant factor that limits the PET image quality. To improve image quality, surveillance systems have been developed to track the movements of the subject during scanning. Gating techniques utilizing the tracking information, are able to compensate for subject motion, thereby improving lesion detection. In this paper, we present a gating method that utilizes the geometric sensitivity gating (GSG) of a 3D-PET scanner system operating in list event acquisition mode. GSG method has several advantages over the existing methods, it only uses LOR events and is non-invasive, no additional hardware device is required and there is no additional patient prepar..

View full abstract