Acoustic parameters measured by a formant-estimating speech processor for a multiple-channel cochlear implant

PJ Blamey, RC Dowell, GM Clark, PM Seligman

Journal of the Acoustical Society of America | ACOUSTICAL SOC AMER AMER INST PHYSICS | Published : 1987

Abstract

In order to assess the limitations imposed on a cochlear implant system by a wearable speech processor, the parameters extracted from a set of 11 vowels and 24 consonants were examined. An estimate of the fundamental frequency EF 0 was derived from the zero crossings of the low-pass filtered envelope of the waveform. Estimates of the first and second formant frequencies EF 1 and EF 2 were derived from the zero crossings of the waveform, which was filtered in the ranges 300-1000 and 800-4000 Hz. Estimates of the formant amplitudes EA 1 and EA 2 were derived by peak detectors operating on the outputs of the same filters. For vowels, these parameters corresponded well to the first and second fo..