Journal article
Capillary whole blood monitoring of oral anticoagulants in children in outpatient clinics and the home setting
V Marzinotto, P Monagle, A Chan, M Adams, P Massicotte, M Leaker, M Andrew
Pediatric Cardiology | Published : 2000
Abstract
A whole blood prothrombin time/international normalized ratio (PT/INR) monitor (CoaguChek, Roche Diagnostics Corp., Indianapolis, IN) was assessed in children for its accuracy, reliability, safety, and acceptance by health care personnel and patient's families. The PT/INR values measured by the CoaguChek monitor showed an excellent correlation with PT/INR values measured by the Hospital for Sick Children (HSC) laboratory (r = 0.96) and Hamilton Civic Hospitals Research Centre (HCHRC) laboratory (r = 0.92) in clinic patients and a close correlation with PT/INR values measured by the HSC laboratory (r = 0.76) and HCHRC laboratory (r = 0.74) in patients at home. Reduced correlation in the home ..
View full abstract