Journal article
Long term OSLD reader stability in the ACDS level one audit
ADC Alves, J Lye, J Kenny, L Dunn, J Lehmann, A Cole, T Kron, D Butler, P Johnston, I Williams
Australasian Physical and Engineering Sciences in Medicine | SPRINGER | Published : 2015
Open access
Abstract
The Australian Clinical Dosimetry Service (ACDS) has demonstrated the capacity to perform a basic dosimetry audit on all radiotherapy clinics across Australia. During the ACDS’s three and a half year trial the majority of the audits were performed using optically stimulated luminescence dosimeters (OSLD) mailed to facilities for exposure to a reference dose, and then returned to the ACDS for analysis. This technical note investigates the stability of the readout process under the large workload of the national dosimetry audit. The OSLD readout uncertainty contributes to the uncertainty of several terms of the dose calculation equation and is a major source of uncertainty in the audit. The st..