Journal article
Stereotactic fields shaped with a micro-multileaf collimator: Systematic characterization of peripheral dose
ML Taylor, LN McDermott, PN Johnston, M Haynes, T Ackerly, T Kron, RD Franich
Physics in Medicine and Biology | IOP PUBLISHING LTD | Published : 2010
Abstract
Despite the highly localized doses that may be delivered via stereotactic radiotherapy, a small dose is nonetheless delivered to out-of-field regions, which may cause detriment to the patient. In this work, a systematic set of dose measurements have been undertaken up to a distance of 45 cm from the isocentre, for stereotactic fields shaped by a BrainLAB mini-multileaf collimator (MMLC) mounted on a Varian 600C linear accelerator. A range of treatment parameters were varied so as to determine the factors of greatest influence and establish relationships with dose. The commercial treatment planning software (TPS) miscalculates the dose to out-of-field regions. Measured dose decreases consiste..
View full abstractGrants
Awarded by Australian Research Council (ARC)
Awarded by Australian Research Council
Funding Acknowledgements
The authors wish to thank C Lancaster (WBRC, The Alfred, Melbourne) for his assistance with dose measurements. This work was supported by an Australian Research Council (ARC) Linkage Project (grant number LP0562315).