Journal article
Flow-injection approach for the determination of the dynamic response characteristics of ion-selective electrodes. Part 2. Application to tubular solid-state iodide electrode
K Tóth, E Lindner, E Pungor, SD Kolev
Analytica Chimica Acta | ELSEVIER SCIENCE BV | Published : 1990
Abstract
The flow-injection approach proposed in Part 1 for investigating the dynamic behaviour of flow-through detectors was applied to the study of a tubular iodide electrode prepared from a pressed precipitate of silver iodide. It was found that the electrode response cannot be described by either of the models developed in Part 1 on the basis of the concepts in the literature for the transient response mechanism of ion-selective electrodes. Scratches and microcavities in the electrode bore were observed under the microscope, most probably resulting from bore drilling. These surface non-idealities make the accessibility of the electrode surface to the analyte variable. A mathematical model develop..
View full abstract