Measurement area and repeatability of semiautomated assessment of corneal endothelium in the topcon specular microscope SP-2000P and IMAGEnet system

X Ding, Q Huang, Y Zheng, Y Jiang, S Huang, M He

Cornea | LIPPINCOTT WILLIAMS & WILKINS | Published : 2012

Abstract

PURPOSE: To investigate the repeatability of the semiautomatic assessment of corneal endothelial cells and its association with the measurement area in the Topcon SP-2000P microscope and IMAGEnet system. METHODS: Specular microscopic images of 86 healthy subjects were captured and analyzed using the Topcon SP-2000P microscope and IMAGEnet system. The same images were analyzed twice, on separate days, by the same examiner using the built-in measurement tool of the IMAGEnet system. The measurement areas were defined with a frame mounted on a computer screen. Four different-sized measurement areas were chosen for the semiautomatic measurements: box A (5.4 × 13.9 cm), box B (4 × 10 cm), box C (4..