Journal article
Continuous measurement of arterial and end-tidal carbon dioxide during cardiac surgery: P(a-ET)CO2 gradient
PS Myles, DA Story, MA Higgs, MR Buckland
Anaesthesia and Intensive Care | AUSTRALIAN SOC ANAESTHETISTS | Published : 1997
Abstract
There have been reports of a negative arterial to end-tidal CO2 gradient (P(a-ET)CO2) during cardiac surgery, so we used capnometry and an intravascular blood gas sensor (Paratrend 7) to continuously monitor this gradient in 20 cardiac surgical patients. We also compared the values obtained from this sensor with those obtained from a standard blood gas analyser at seven time points. We found a significant change in P(a-ET)CO2 after cardopulmonary bypass (P < 0.001) though we were unable to demonstrate a negative P(a-ET)CO2 at any time (95% CI 0.14%). There was clinically acceptable agreement between labarotary and Paratrend 7 measurements during and after cardiac surgery.