The influence of acquisition mode on the dosimetric performance of an amorphous silicon electronic portal imaging device
O Bawazeer, S Herath, S Sarasanandarajah, T Kron, P Deb
Journal of Medical Physics | MEDKNOW PUBLICATIONS & MEDIA PVT LTD | Published : 2017
Abstract
Aims: This study investigates the impact of cine acquisition mode on the dosimetric characteristics of a Varian aS500 amorphous silicon electronic portal imaging device (a-Si EPID). Materials and Methods: The performance of an a-Si EPID operated in cine mode was assessed and compared to its performance when operated in an integrated mode and dose measurements using an ionization chamber. This study was conducted at different photon energies and the EPID performance was assessed as function of the delivered dose, dose rate, multileaf collimator speed, field size, phantom thickness, and intensity-modulated radiation therapy fields. Results: The worst nonlinearity was observed at low monitor un..