Journal article

Stereotactic radiosurgery planning with ictal SPECT images.

T Ackerly, M Geso, G O'Keefe, R Smith

Physical and Engineering Sciences in Medicine | Published : 2004

Abstract

This paper is motivated by a clinical requirement to utilise ictal SPECT images for target localisation in stereotactic radiosurgery treatment planning using the xknife system which only supports CT and MRI images. To achieve this, the SPECT images were converted from raw (pixel data only) format into a part 10 compliant DICOM CT fileset. The minimum requirements for the recasting of a raw format image as DICOM CT or MRI data set are described in detail. The method can be applied to the importation of raw format images into any radiotherapy treatment planning system that supports CT or MRI import. It is demonstrated that the combination of the low spatial resolution SPECT images, depicting f..

View full abstract

University of Melbourne Researchers