Objectives This study sought to validate a 3-dimensional (3D) phase mapping system and determine the distribution of dominant propagation patterns in persistent atrial fibrillation (AF). Background Currently available systems display phase as simplified 2-dimensional maps. We developed a novel 3D phase mapping system that uses the 3D location of basket catheter electrodes and the patient's 3D left atrial surface geometry to interpolate phase and create a 3D representation of phase progression. Methods Six-min AF recordings from the left atrium were obtained in 14 patients using the Constellation basket catheter and analyzed offline. Exported signals underwent both phase and traditional activ..