Journal article

Minimally invasive cardiac output monitoring: agreement of oesophageal Doppler, LiDCOrapid™ and Vigileo FloTrac™ monitors in non-cardiac surgery.

TD Phan, R Kluger, C Wan

Anaesthesia and Intensive Care | Published : 2016

Abstract

There is lack of data about the agreement of minimally invasive cardiac output monitors, which make it impossible to determine if they are interchangeable or differ objectively in tracking physiological trends. We studied three commonly used devices: the oesophageal Doppler and two arterial pressure-based devices, the Vigileo FloTrac™ and LiDCOrapid™. The aim of this study was to compare the agreement of these three monitors in adult patients undergoing elective non-cardiac surgery. Measurements were taken at baseline and after predefined clinical interventions of fluid, metaraminol or ephedrine bolus. From 24 patients, 131 events, averaging 5.2 events per patient, were analysed. The cardiac..