Journal article
Six-month Longitudinal Comparison of a Portable Tablet Perimeter With the Humphrey Field Analyzer
SM Prea, YXG Kong, A Mehta, M He, JG Crowston, V Gupta, KR Martin, AJ Vingrys
American Journal of Ophthalmology | ELSEVIER SCIENCE INC | Published : 2018
Abstract
Purpose: To establish the medium-term repeatability of the iPad perimetry app Melbourne Rapid Fields (MRF) compared to Humphrey Field Analyzer (HFA) 24-2 SITA-standard and SITA-fast programs. Design: Multicenter longitudinal observational clinical study. Methods: Sixty patients (stable glaucoma/ocular hypertension/glaucoma suspects) were recruited into a 6-month longitudinal clinical study with visits planned at baseline and at 2, 4, and 6 months. At each visit patients undertook visual field assessment using the MRF perimetry application and either HFA SITA-fast (n = 21) or SITA-standard (n = 39). The primary outcome measure was the association and repeatability of mean deviation (MD) for t..
View full abstractGrants
Funding Acknowledgements
CAMBRIDGE EYE TRUST, CAMBRIDGE UK, JUKES GLAUCOMA RESEARCH FUND UK, GLANCE OPTICAL PTY. Ltd., Melbourne, Australia,