Strategies for enumeration of circulating microvesicles on a conventional flow cytometer: Counting beads and scatter parameters
MJ Alkhatatbeh, AK Enjeti, S Baqar, EI Ekinci, D Liu, RF Thorne, LF Lincz
Journal of Circulating Biomarkers | Published : 2018
Abstract
Enumeration of circulating microvesicles (MVs) by conventional flow cytometry is accomplished by the addition of a known amount of counting beads and calculated from the formula: MV/μl = (MV count/bead count) × final bead concentration. We sought to optimize each variable in the equation by determining the best parameters for detecting ‘MV count’ and examining the effects of different bead preparations and concentrations on the final calculation. Three commercially available bead preparations (TruCount, Flow-Count and CountBright) were tested, and MV detection on a BD FACSCanto was optimized for gating by either forward scatter (FSC) or side scatter (SSC); the results were compared by calcul..