Journal article

Clinical performance of a novel textile interface for neonatal chest electrical impedance tomography

L Sophocleous, I Frerichs, M Miedema, M Kallio, T Papadouri, C Karaoli, T Becher, DG Tingay, AH Van Kaam, R Bayford, AD Waldmann

Physiological Measurement | IOP PUBLISHING LTD | Published : 2018

Abstract

Objective: Critically ill neonates and infants might particularly benefit from continuous chest electrical impedance tomography (EIT) monitoring at the bedside. In this study a textile 32-electrode interface for neonatal EIT examination has been developed and tested to validate its clinical performance. The objectives were to assess ease of use in a clinical setting, stability of contact impedance at the electrode-skin interface and possible adverse effects. Approach: Thirty preterm infants (gestational age: 30.3 ± 3.9 week (mean ± SD), postnatal age: 13.8 ± 28.2 d, body weight at inclusion: 1727 ± 869 g) were included in this multicentre study. The electrode-skin contact impedances were mea..

View full abstract

University of Melbourne Researchers