Journal article

Smart watches for heart rate assessment in atrial arrhythmias

AN Koshy, JK Sajeev, N Nerlekar, AJ Brown, K Rajakariar, M Zureik, MC Wong, L Roberts, M Street, J Cooke, AW Teh

International Journal of Cardiology | ELSEVIER IRELAND LTD | Published : 2018

Abstract

Background: Despite studies demonstrating the accuracy of smart watches (SW) and wearable heart rate (HR) monitors in sinus rhythm, no data exists regarding their utility in arrhythmias. Methods: 102 hospitalized patients were evaluated at rest using continuous electrocardiogram (ECG) monitoring with concomitant SW-HR (FitBit, FB, Apple Watch, AW) for 30 min. Results: Across all devices, 38,616 HR values were recorded. Sinus rhythm cohort demonstrated strong agreement for both devices with a low bias (FB & AW Bias = 1 beat). In atrial arrhythmias, AW demonstrated a stronger correlation than FB (AW rs = 0.83, FB rs = 0.56, both p < 0.01) with a lower bias (Bias AW = −5 beats, FB = −18 beats)...

View full abstract

Grants

Awarded by Eastern Health Foundation


Funding Acknowledgements

This work was supported by the Eastern Health Foundation Research Grant [EHFRG2017_029]. The sponsor had no role in study design, collection, analysis, interpretation of data and in the decision to submit the article for publication. Dr. Nerlekar is supported by the National Health and Medical Research Council of Australia and National Heart Foundation Scholarship (1115172). Dr. Teh is supported by an Early Career Fellowship from the National Health and Medical Research Council of Australia.