Journal article

Sensitivity of electronic portal imaging device (EPID) based transit dosimetry to detect inter-fraction patient variations

O Bawazeer, S Sarasanandarajah, S Herath, T Kron, P Deb

Ifmbe Proceedings | SPRINGER | Published : 2019

Abstract

The sensitivity of EPID-based transit dosimetry to detect patient variations between treatment fractions is examined using gamma analysis and a structural similarity (SSIM) index. EPID images were acquired for 3-dimensional conformal (3DCRT) and dynamic intensity modulated (dIMRT) radiation therapy fields in multiple fractions. Transit images were converted to doses, transit dose in the first fraction considered the reference dose. Variations in patient position or weight were then introduced in the subsequent fractions. Positional variations were examined using a lung and a head and neck phantoms. Anatomical variations were examined using a slab phantom in three scenarios, with solid water ..

View full abstract

University of Melbourne Researchers