Journal article
Pulse Oximeter Saturation Targeting and Oximeter Changes in the Benefits of Oxygen Saturation Targeting (BOOST)-II Australia and BOOST-II UK Oxygen Trials
BJ Stenson, M Donoghoe, P Brocklehurst, PG Davis, E Juszczak, IC Marschner, J Simes, WO Tarnow-Mordi
Journal of Pediatrics | MOSBY-ELSEVIER | Published : 2019
Abstract
Infants in the Australian and UK Benefits of Oxygen Saturation Targeting-II trials treated using revised oximeters spent more time within their planned pulse oximeter saturation target ranges than infants treated using the original oximeters (P <.001). This may explain the larger mortality difference seen with revised oximeters. If so, average treatment effects from the Neonatal Oxygen Prospective Meta-analysis trials may be underestimates.
Grants
Awarded by Medical Research Council