Journal article
Scalar localisation of peri-modiolar electrodes and speech perception outcomes
C Shaul, AS Dragovic, AK Stringer, SJ O'Leary, RJ Briggs
Journal of Laryngology and Otology | CAMBRIDGE UNIV PRESS | Published : 2018
Open access
Abstract
Objective To identify the intracochlear electrode position in cochlear implant recipients and determine the correlation to speech perception for two peri-modiolar electrode arrays.Methods Post-operative cone-beam computed tomography images of 92 adult recipients of the 'CI512' electrode and 18 adult recipients of the 'CI532' electrode were analysed. Phonemes scores were recorded pre-implantation, and at 3 and 12 months post-implantation.Results All CI532 electrodes were wholly within scala tympani. Of the 79 CI512 electrodes intended to be in scala tympani, 58 (73 per cent) were in scala tympani, 14 (17 per cent) were translocated and 7 (9 per cent) were wholly in scala vestibuli. Thirteen C..
View full abstract