Journal article
Dosimetric end-to-end tests in a national audit of 3D conformal radiotherapy
J Lehmann, A Alves, L Dunn, M Shaw, J Kenny, S Keehan, J Supple, F Gibbons, S Manktelow, C Oliver, T Kron, I Williams, J Lye
Physics and Imaging in Radiation Oncology | ELSEVIER | Published : 2018
Abstract
Background and purpose: Independent dosimetry audits improve quality and safety of radiation therapy. This work reports on design and findings of a comprehensive 3D conformal radiotherapy (3D-CRT) Level III audit. Materials and methods: The audit was conducted as onsite audit using an anthropomorphic thorax phantom in an end-to-end test by the Australian Clinical Dosimetry Service (ACDS). Absolute dose point measurements were performed with Farmer-type ionization chambers. The audited treatment plans included open and half blocked fields, wedges and lung inhomogeneities. Audit results were determined as Pass Optimal Level (deviations within 3.3%), Pass Action Level (greater than 3.3% but wit..
View full abstract