Journal article

A robust VMAT delivery solution for single-fraction lung SABR utilizing FFF beams minimizing dosimetric compromise

A Burton, K Offer, N Hardcastle

Journal of Applied Clinical Medical Physics | WILEY | Published : 2020

Open access

Abstract

Peripheral lung lesions treated with a single fraction of stereotactic ablative body radiotherapy (SABR) utilizing volumetric modulated arc therapy (VMAT) delivery and flattening filter-free (FFF) beams represent a potentially high-risk scenario for clinically significant dose blurring effects due to interplay between the respiratory motion of the lesion and dynamic multi-leaf collimators (MLCs). The aim of this study was to determine an efficient means of developing low-modulation VMAT plans in the Eclipse treatment planning system (v15.5, Varian Medical Systems, Palo Alto, USA) in order to minimize this risk, while maintaining dosimetric quality. The study involved 19 patients where an int..

View full abstract

University of Melbourne Researchers