Journal article

Does variability in automated perfusion software outputs for acute ischemic stroke matter? Reanalysis of EXTEND perfusion imaging

A Bivard, L Churilov, H Ma, C Levi, B Campbell, N Yassi, A Meretoja, H Zhao, G Sharma, C Chen, S Davis, G Donnan, B Yan, M Parsons

CNS Neuroscience and Therapeutics | WILEY | Published : 2022

Open access

Abstract

Aims: We reprocessed the Extending the time for Thrombolysis in Emergency Neurological Deficits (EXTEND) perfusion imaging with a different automated software with the aim of comparing mismatch eligibility and outcomes. Methods: EXTEND baseline perfusion imaging data were reprocessed using autoMIStar software to identify patients who were eligible based on the same target mismatch criteria as per the original trial. Results: From the 225 patients fulfilling RAPID-based mismatch criteria randomized in the EXTEND study, 196 (87%) patients met the revised mismatch criteria. Most common reasons for not meeting revised criteria were core >70 ml (n = 9), and no perfusion lesion/lack of penumbral t..