Journal article
The determination of zinc using flow injection and continuous flow analysis combined with a polymer inclusion film-coated column: Application to the determination of zinc in alloys and commercial lithium chloride
S Kagaya, K Hida-Matsuda, S Tsuzaka, C Minami, M Gemmei-Ide, RW Cattrall, SD Kolev
Talanta | Published : 2023
Abstract
A column coated with a polymer inclusion film (PIF) containing Aliquat 336 as carrier cast on glass beads packed in a glass tube is described for the separation, preconcentration, and determination of zinc(II) in flow injection analysis (FIA) and continuous flow analysis (CFA) systems. In the FIA method, 200 μL of a sample solution containing 2 mol/L lithium chloride is injected into a 2 mol/L lithium chloride stream. This converts zinc(II) ion into its anionic chlorocomplexes which are then extracted into the Aliquat 336-based PIF by anion exchange. The extracted zinc(II) is then back-extracted into a stream of 1 mol/L sodium nitrate solution and determined spectrophotometrically using 4-(2..
View full abstract