Journal article

The detection of ACTH and insulin in equine plasma by solid-phase extraction and micro-flow LC/MSMS

R Steel, M Timms, N Bamford, R Spence, M Sillence

Drug Testing and Analysis | WILEY | Published : 2025

Abstract

Previous liquid chromatography/mass spectrometry (LC/MS) methods for the detection of insulin and other similar peptide hormones in equine plasma relied on the use of antibody affinity extraction. As a result, these methods were not suitable for routine high-throughput analysis. A solid-phase extraction (SPE) method incorporating size exclusion as well as reversed-phase interactions allows the selective extraction of peptide hormones such as adrenocorticotropic hormone (ACTH), insulin and their synthetic analogues from equine plasma with approximately 80% extraction efficiencies. This extraction was combined with on-column derivatisation with acetic anhydride, followed by tryptic digestion a..

View full abstract

University of Melbourne Researchers