Journal article

Quantification of the difference in hounsfield units of an electron density phantom between a conventional and standing computed tomography machine

C Beck, AV Morrice-West, P Muir, PL Hitchens, RC Whitton

Veterinary Research Communications | Published : 2025

Abstract

In human CT imaging, opportunistic calculation of bone mineral density (BMD) is used to screen for osteoporosis, where BMD is estimated by Hounsfield Units (HU). There is potential to use HU for estimation of BMD in veterinary imaging. However, HU values can vary between different machines and over time. We undertook a serial measurements study to determine (1) the extent of the difference of the mean HU and noise (HU standard deviation) of eight regions of interest (ROI) of an electron density phantom scanned on two CT machines, a Siemens CT (conventional) machine and the Equina by Asto CT (standing) machine; and (2) to determine if the HU values obtained on each machine differed over time...

View full abstract