Journal article

External quality assessment program for Chlamydia trachomatis diagnostic testing by nucleic acid amplification assays

S Land, S Tabrizi, A Gust, E Johnson, S Garland, EM Dax

Journal of Clinical Microbiology | AMER SOC MICROBIOLOGY | Published : 2002

Abstract

We report the results from 57 Australian diagnostic laboratories testing two external quality assessment panels using either the Roche Amplicor Chlamydia trachomatis test (R-PCR) or the Abbott LCx Chlamydia trachomatis assay (A-ligase chain reaction [LCR]). Panel samples were either normal urine spiked with Chlamydia trachomatis antigen or clinical urine specimens. There was no significant difference between laboratories or between assays in detection of C. trachomatis-positive clinical samples. Only at the lower limit of detection of the assays did the R-PCR demonstrate increased sensitivity over the A-LCR in the detection of C. trachomatis antigen. However, it was found that single-sample ..