Journal article
Ion-beam-induced-charge characterisation of particle detectors
C Yang, DN Jamieson, SM Hearne, CI Pakes, B Rout, E Gauja, AJ Dzurak, RG Clark
Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms | ELSEVIER SCIENCE BV | Published : 2002
Abstract
Ion-beam-induced-charge collection (IBIC) in a nuclear microprobe has been used to characterise detectors for the measurement of particles over a median energy range (100 keV-1 MeV). Three standard detector devices have been studied: a PIPS detector with a buried (ion-implanted) junction structure, a Schottky barrier junction device and a PN-junction photodiode. A 2.0 MeV focussed helium ion beam was used to probe the active area of each device with a spatial resolution ∼1-2 μm, to quantify the thickness of the dead layer, the charge collection response and the reduction in charge collection efficiency induced by ion-beam damage. © 2002 Elsevier Science B.V. All rights reserved.