Journal article

Performance of efficient test procedures for frequency-doubling technology perimetry in normal and glaucomatous eyes

A Turpin, AM McKendrick, CA Johnson, AJ Vingrys

Investigative Ophthalmology and Visual Science | ASSOC RESEARCH VISION OPHTHALMOLOGY INC | Published : 2002

Abstract

PURPOSE. To validate the clinical performance of two new efficient threshold-estimation procedures for frequency-doubling technology (FDT) perimetry in both visually normal individuals and patients with glaucomatous visual field loss. METHODS. Forty-one normal subjects (mean age, 48.3 ± 11.6 years) and 50 patients with glaucomatous visual field loss (mean age, 72.7 ± 10.0 years) were tested. Some of these participants were retested within a 3-month period. FDT perimetry was performed on a color monitor driven by a visualstimulus-generating video board, with stimulus parameters designed to closely mimic those of the commercial FDT test. Visual field sensitivity was measured using three proced..

View full abstract

University of Melbourne Researchers