Abstract
Two-dimensional backgammon configuration and multiwire gas proportional counters have been in use for decades in the field of x-ray physics for high-precision or high-flux experiments. Systematics inherent to this type of detector lead to image distortions that are often overlooked or extremely difficult to quantify. They are usually seen as intrinsic to the detector performance, leading to loss of resolution, linearity and signal-to-noise ratio. This work presents the signature, cause and resolution of several key distortions. The physics of the breakdown between the anode and cathode, false event position reconstruction due to the electronic detector response, and the effect of asymmetric ..