Journal article

SU‐FF‐I‐66: MRI Relaxometry BMD Measurements Using Conventional Phase Symmetrized Rapid Increased Flip Spin Echo and Standard Gradient Echo and Its Correlation with DXA

M Bakhtiary, N Riyahi‐Alami, M Oghabian, A Ghasemzadeh, H Ghanaati, S Sarkar

Medical Physics | Wiley | Published : 2006

Abstract

Purpose: Recently, several special MRI protocols have been developed for BMD measurements for deriving R2(=1/T2), R2*(=1/T2*) and R2′(=R2*‐R2) applied in several cross sectional studies. But, due to high slew rate gradients with high amplitudes in small periods, such protocols can be used only in system with high slew rate (22 mT/m.s or greater). However, our main aim in this study was to develop optimized versions of standard GE and conventional SE protocols using in existing MRI systems for BMD measurements, as most systems do not support such newly high slew rate protocols. Method and Materials: This study was performed by 1.5T‐MRI system(Picker Vista‐Q800, slew rate=13mT/m.s), SNR phanto..

View full abstract

University of Melbourne Researchers