Journal article

Data acquisition system development for the detection of X-ray photons in multi-wire gas proportional counters

JA Kimpton, MN Kinnane, LF Smale, CT Chantler, LT Hudson, A Henins, CI Szabo, JD Gillaspy, JN Tan, JM Pomeroy, E Takacs, B Radics

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment | ELSEVIER | Published : 2007

Abstract

A new data acquisition system coupled to a backgammon-type gas proportional counter capable of single-photon counting over a wide range of count rates has been developed and replaces a CAMAC-based system. The new apparatus possesses improved architecture, interface technology, speed and diagnostic capability. System efficiency and throughput is significantly improved, especially in addressing earlier problems of hardware buffer downloads containing zero or repeat data and inefficient gating control. The new system is a PXI-based data acquisition apparatus including additional electronics, controlled by a graphical programming environment. It allows development of superior diagnostic tools fo..

View full abstract

University of Melbourne Researchers