Effect of vacuum profile on breast milk expression using an electric breast pump.

Leon R Mitoulas, Ching Tat Lai, Lyle C Gurrin, Michael Larsson, Peter E Hartmann

Journal of Human Lactation | Published : 2002

Abstract

The authors compared milk expression using 5 experimental vacuum patterns and a commercially available vacuum pattern ranging in cycle times (20 to 78 cycles/min) and vacuum curve dynamics in 30 mothers using an experimental, software-controlled electric breast pump. The volume of milk removed over 5 minutes differed (P = .0072) between patterns (range = 62.8 +/- 6.6 mL to 47.2 +/- 5.1 mL). However, there was no difference in the percentage of available milk removed (range = 99.4% +/- 15.1% to 70.6% +/- 8.6%). The rate of milk removal differed between patterns at both the beginning (1 minute) and end (1.5 minutes) of the expression period (P < .05). Peak vacuum chosen differed between patter..