Abstract

Delivering power to an implanted device located deep inside the body is not trivial. This problem is made more challenging if the implanted device is in constant motion. This paper describes two methods of transferring power wirelessly by means of magnetic induction coupling. In the first method, a pair of transmit and receive coils is used for power transfer over a large distance (compared to their diameter). In the second method, an intermediate pair of coils is inserted in between transmit and receive coils. Comparison between the power transfer efficiency with and without the intermediate coils shows power transfer efficiency to be 11.5 % and 8.8 %, respectively. The latter method is esp..