Journal article
Integrated oxygenator FX05
SB Horton, S Donath, CA Thuys, MJ Bennett, SL Augustin, AM Horton, BJ Schultz, SJ Bottrell, I Konstantinov, Y D'Udekem, C Brizard
ASAIO Journal | LIPPINCOTT WILLIAMS & WILKINS | Published : 2011
Abstract
In recent years, there has been a trend to incorporate arterial filters into cardiopulmonary bypass (CPB) oxygenators. The FX oxygenators are the first examples of integrating a filter around the fibers of the oxygenator. To assess the efficacy of the FX05 oxygenator, in comparison with our existing CPB setup (RX05 oxygenator + arterial filter), we randomly assigned 40 patients undergoing CPB, with calculated flows <1.5 L/min, to FX05 or RX05 + arterial filter (19 FX and 21 RX). Embolic load was assessed using the emboli detection and classification quantifier, with sensors placed prefilter and postfilter/oxygenator at identical distances for both devices. A total of 6,270 FX readings and 6,..
View full abstractGrants
Funding Acknowledgements
This study was partially supported by the Murdoch Childrens Research Institute.