Journal article
Oral continuous positive airway pressure (CPAP) following nasal injury in a preterm infant
HR Carlisle, COF Kamlin, LS Owen, PG Davis, CJ Morley
Archives of Disease in Childhood Fetal and Neonatal Edition | B M J PUBLISHING GROUP | Published : 2010
Abstract
Non-invasive respiratory support is increasingly popular but is associated with complications including nasal trauma. The present report describes a novel method of oral continuous positive airway pressure (CPAP) delivery in an extremely premature infant with severe nasal septum erosion. The distal end of a cut down endotracheal tube was passed through a small hole made in the teat of a dummy (infant pacifi er) and sutured in place. The dummy was secured in the infant's mouth and CPAP was delivered to the pharynx. The device was well tolerated and the infant was successfully managed using this technique for 48 days, avoiding endotracheal intubation and ventilation.