Journal article
End tidal carbon dioxide is as reliable as transcutaneous monitoring in ventilated postsurgical neonates.
DG Tingay, KS Mun, EJ Perkins
Archives of Disease in Childhood Fetal and Neonatal Edition | Published : 2013
Abstract
To compare the agreement, precision and repeatability of end tidal carbon dioxide (EtCO2 ) and transcutaneous carbon dioxide (TcCO2 ) with partial pressure of arterial CO(2) ( PaCO2) in postoperative neonates. Fifty mechanically ventilated neonates without lung disease, and with no contraindications for either TcCO2 or EtCO2 monitoring. Paired TcCO2 and EtCO2 values were recorded with three consecutive measurements within the first 48 h of surgery. EtCO2, TcCO2 and PaCO2 triplets were compared using Bland-Altman plots. One hundred thirty-two triplet measures of CO(2) were recorded with mean PaCO2 43.5 (7.3) mm Hg, EtCO2 38.8 (6.4) mm Hg and 43.8 (8.8) mm Hg (p<0.0001 for EtCO2 against PaCO2;..
View full abstractGrants
Awarded by National Health and Medical Research Council Clinical Research Fellowship
Funding Acknowledgements
DGT is supported by a National Health and Medical Research Council Clinical Research Fellowship (Grant ID 491286) and the Victorian Government Operational Infrastructure Support Programme.