Abstract
A 240-μm diameter ultrasonic micromotor is presented as a potential solution for an especially difficult task in minimally invasive neurosurgery, navigating a guidewire to an injury in the neurovasculature as the first step of surgery. The peak no-load angular velocity and maximum torque were 600 rad/s and 1.6 nN-m, respectively, and we obtained rotation about all three axes. By using a burst drive scheme, open-loop position and speed control were achieved. The construction method and control scheme proposed in this study remove most of the current limitations in minimally invasive, catheter-based actuation, enabling minimally invasive vascular surgery concepts to be pursued for a broad vari..
View full abstractGrants
Awarded by National Health and Medical Research Council
Funding Acknowledgements
The authors wish to acknowledge the support of this work by funding provided to them by the NHMRC through Development Grant No. 546238 and the CASS Foundation through Grant No. SM/07/1616.