Journal article

Position resolution and particle identification with the ATLAS em calorimeter

J Colas, L Di Ciaccio, M El Kacimi, O Gaumer, M Gouanère, D Goujdami, R Lafaye, C Le Maner, L Neukermans, P Perrodo, L Poggioli, D Prieur, H Przysiezniak, G Sauvage, I Wingerter-Seez, R Zitoun, F Lanni, H Ma, S Rajagopalan, S Rescia Show all

Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment | ELSEVIER | Published : 2005

Abstract

In the years between 2000 and 2002 several pre-series and series modules of the ATLAS EM barrel and end-cap calorimeter were exposed to electron, photon and pion beams. The performance of the calorimeter with respect to its finely segmented first sampling has been studied. The polar angle resolution has been found to be in the range 50-60 (mrad)/E(GeV). The π0 rejection has been measured to be about 3.5 for 90% photon selection efficiency at pT=50GeV/c. e-π separation studies have indicated that a pion fake rate of (0.07-0.5)% can be achieved while maintaining 90% electron identification efficiency for energies up to 40 GeV. © 2005 Elsevier B.V. All rights reserved.

University of Melbourne Researchers