Test-retest reliability of retinal oxygen saturation measurement

RA O'Connell, AJ Anderson, SL Hosking, AH Batcha, BV Bui

Optometry and Vision Science | LIPPINCOTT WILLIAMS & WILKINS | Published : 2014

Abstract

PURPOSE: To determine intrasession and intersession repeatability of retinal vessel oxygen saturation from the Oxymap Retinal Oximeter using a whole image-based analysis technique and so determine optimal analysis parameters to reduce variability. METHODS: Ten fundus oximetry images were acquired through dilated pupils from 18 healthy participants (aged 22 to 38) using the Oxymap Retinal Oximeter T1. A further 10 images were obtained 1 to 2 weeks later from each individual. Analysis was undertaken for subsets of images to determine the number of images needed to return a stable coefficient of variation (CoV). Intrasession and intersession variability were quantified by evaluating the CoV and..