Journal article

Comparison of TLD calibration methods for 192Ir dosimetry

A Haworth, DJ Butler, L Wilfert, MA Ebert, SP Todd, AJM Hayton, T Kron

Journal of Applied Clinical Medical Physics | MULTIMED INC | Published : 2013

Abstract

For the purpose of dose measurement using a high-dose rate 192Ir source, four methods of thermoluminescent dosimeter (TLD) calibration were investigated. Three of the four calibration methods used the 192Ir source. Dwell times were calculated to deliver 1 Gy to the TLDs irradiated either in air or water. Dwell time calculations were confirmed by direct measurement using an ionization chamber. The fourth method of calibration used 6 MV photons from a medical linear accelerator, and an energy correction factor was applied to account for the difference in sensitivity of the TLDs in 192Ir and 6 MV. The results of the four TLD calibration methods are presented in terms of the results of a brachyt..

View full abstract