Journal article

Note: Calibration of atomic force microscope cantilevers using only their resonant frequency and quality factor

John E Sader, James R Friend

REVIEW OF SCIENTIFIC INSTRUMENTS | AMER INST PHYSICS | Published : 2014

Abstract

A simplified method for calibrating atomic force microscope cantilevers was recently proposed by Sader et al. [Rev. Sci. Instrum. 83, 103705 (2012); Sec. III D] that relies solely on the resonant frequency and quality factor of the cantilever in fluid (typically air). This method eliminates the need to measure the hydrodynamic function of the cantilever, which can be time consuming given the wide range of cantilevers now available. Using laser Doppler vibrometry, we rigorously assess the accuracy of this method for a series of commercially available cantilevers and explore its performance under non-ideal conditions. This shows that the simplified method is highly accurate and can be easily i..

View full abstract

University of Melbourne Researchers

Grants

Funding Acknowledgements