Journal article

Determination of lamotrigine in human plasma by high-performance liquid chromatography

P Angelis-Stoforidis, DJ Morgan, TJ O'Brien, FJE Vajda

Journal of Chromatography B Biomedical Sciences and Applications | ELSEVIER SCIENCE BV | Published : 1999

Abstract

The method involves precipitation of plasma proteins with acetonitrile and analysis of the supernatant by high-performance liquid chromatography using a 5 μm Zorbax C8 column. Quantitation was performed by measurement of the UV absorbance at a wavelength of 306 nm. The method was linear in the range of 1-20 μg/ml, with a mean coefficient of determination (r2=0.998). The limit of detection was 0.6 μg/ml and the lower limit of quantitation was 1 μg/ml using 200 μl of plasma. Within- and between-day accuracy and precision were below 6% at all analysed concentrations except at the limit of quantitation. No interfering peaks were found by commonly monitored antiepileptic drugs. Recovery was found..

View full abstract