Evaluation of dosimetric misrepresentations from 3D conventional planning of liver SBRT using 4D deformable dose integration

UA Yeo, ML Taylor, JR Supple, S Siva, T Kron, D Pham, RD Franich

Journal of Applied Clinical Medical Physics | Published : 2014

Open access

Abstract

The purpose of this study is to evaluate dosimetric errors in 3D conventional planning of stereotactic body radiotherapy (SBRT) by using a 4D deformable image registration (DIR)-based dose-warping and integration technique. Respiratory-correlated 4D CT image sets with 10 phases were acquired for four consecutive patients with five liver tumors. Average intensity projection (AIP) images were used to generate 3D conventional plans of SBRT. Quasi-4D path-integrated dose accumulation was performed over all 10 phases using dose-warping techniques based on DIR. This result was compared to the conventional plan in order to evaluate the appropriateness of 3D (static) dose calculations. In addition, ..